Perkin Elmer Lambda 35 family
Research Quality: 30% complete

Provisional entry — research in progress

This page was generated automatically from cited public sources and hasn't completed the full research pass yet. Every specification shown carries its source; more detail is added as research completes.

The Perkin Elmer Lambda 35 is a scanning double-beam UV/Vis spectrophotometer. The Perkin Elmer Lambda 35 has a wavelength range of 190 nm to 1100 nm. The Perkin Elmer Lambda 35 has selectable spectral bandwidths of 0.5 nm, 1 nm, 2 nm, and 4 nm.[1][2]

Perkin ElmerLambda 35
No photo available yet

Power

100 V to 240 V AC, 50/60 Hz; 250 VA[2]

Optics

190 nm to 1100 nm[2]

What it is

The Perkin Elmer Lambda 35 is a UV/Vis spectrophotometer.[1]

What do the numbers mean?

Power & electrical1

Power requirements
100 V to 240 V AC, 50/60 Hz; 250 VA[2]
Accurate?

Optics & imaging24

Wavelength range
190 nm to 1100 nm[2]
Accurate?
Spectral bandwidth
0.5 nm, 1 nm, 2 nm, 4 nm (selectable fixed slit widths)[2]
Accurate?
Detector
Photodiodes, one for the sample beam and one for the reference beam[2]
Accurate?
Photometric accuracy
Absorbance ±0.003 at Absorbance = 1 at 440 nm, 546.1 nm and 635 nm; absorbance ±0.015 at Absorbance = 1 at 257 nm and 350 nm[2]
Accurate?
Stray radiation
Transmission <0.02% at 220 nm, 340 nm and 370 nm; absorbance >2 at 200 nm with potassium chloride solution against distilled water[2]
Accurate?
Baseline stability
Absorbance <0.0003 per hour at 500 nm after warmup[2]
Accurate?
Optical pathlength in sample compartment
121 mm[2]
Accurate?
Modes of operation
Scanning, wavelength program, time-drive, rate, quant, scanning quant[4]
Accurate?
Dimensions
Width: 650 nm; Height: 260 nm; Depth: 560 nm[2]
Accurate?
Wavelength range
190 nm to 1100 nm; 0 nm for alignment purposes[2]
Accurate?
Wavelength accuracy
±0.1 nm[2]
Accurate?
Wavelength reproducibility
±0.1 nm[2]
Accurate?
Lamp change
Automatically at 326 nm (software selectable over the whole wavelength range)[2]
Accurate?
Scan speeds
7.5, 15, 30, 60, 120, 240, 480, 960, 1920 and 2880 nm/min[2]
Accurate?
Photometric accuracy
Absorbance ±0.003 at Absorbance = 1 at 440 nm, 546.1 nm and 635 nm with NIST 930 filters; Absorbance ±0.015 at Absorbance = 1 at 257 nm and 350 nm with potassium dichromate solution[2]
Accurate?
Baseline linearity
2 nm slit: Absorbance ±0.0005 (corrected: 200 nm to 1100 nm, scan speed 240 nm/min, smooth 2)[2]
Accurate?
Baseline noise
2 nm slit: Absorbance < 0.00006 RMS; Absorbance < 0.0002 peak-to-peak 3 min at Absorbance = 0, wavelength 500 nm and response 2 s[2]
Accurate?
Baseline stability
Absorbance < 0.0003 per hour (500 nm, after warmup)[2]
Accurate?
Optical system
All-reflecting optical system with silica-coated optical components and a holographic concave grating with 1053 lines/mm in the center[2]
Accurate?
Beam cross-section
0.5 nm slit ca. 0.25 mm x 7 mm; 1 nm slit ca. 0.5 mm x 7.5 mm; 2 nm slit ca. 1 mm x 7.5 mm; 4 nm slit ca. 2 mm x 7.5 mm[2]
Accurate?
Optical system
All-reflecting optical system with silica-coated optical components and a holographic concave grating[2]
Accurate?
Baseline linearity
Absorbance ±0.0005 for 2 nm slit; corrected 200 nm to 1050 nm, scan speed 240 nm/min, smooth 2[2]
Accurate?
Baseline noise
Absorbance < 0.00006 RMS; Absorbance < 0.0002 peak-to-peak, 3 min at Absorbance = 0, wavelength 500 nm and response 2 s[2]
Accurate?
Baseline stability (drift)
Absorbance < 0.0003 per hour at 500 nm after warmup[2]
Accurate?

Control & software5

Instrument type
Scanning double-beam spectrometer for the UV/Vis range; operation by PC[2]
Accurate?
Interface
One RS 232 C interface (serial) for connecting a PC[2]
Accurate?
Type
Scanning double-beam spectrometer for the UV/Vis range; operation by PC[2]
Accurate?
Data output
One RS 232 C interface (serial), for connecting a PC[2]
Accurate?
Data output
One RS 232 C serial interface for connecting a PC[2]
Accurate?

Dimensions & facilities2

Beam center height
15 mm above cell holder bottom[2]
Accurate?
Mass
26 kg approx.[2]
Accurate?

Configuration & options7

Radiation sources
Prealigned deuterium and halogen lamps[2]
Accurate?
Photometric range
Transmission 0% to 100%; absorbance -6.000 to 6.000; concentration units 1 to 9999[2]
Accurate?
Use cases
Measurements on liquids, solids, pastes and powder samples[3]
Accurate?
Ambient operating temperature
15 ºC to 35 ºC[2]
Accurate?
Humidity range
20% to 80% relative humidity without condensation[2]
Accurate?
Photometric range
Transmission 0% to 100%; Absorbance –6.000 to 6.000 (display range); 1 to 9999 (concentration units)[2]
Accurate?
Grating
Holographic concave grating with 1053 lines/mm in the center[2]
Accurate?

Vintage & configurations

Documented models & variants

DesignationGenerationVintageChangesSource
Lambda 25——Fixed 1 nm spectral bandwidth; Lambda 35 has selectable fixed slit widths of 0.5, 1, 2, or 4 nm.richmondscientific.com[2]
Lambda 45——Lambda 45 substitutes M2 with a pre-monochromator grating, increasing straylight performance; Lambda 45 has selectable fixed slit widths of 0.5, 1, 2, or 4 nm.richmondscientific.com[2]
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Power requirements100 V to 240 V AC, 50/60 Hz; 250 VA[2]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What type of instrument is the Perkin Elmer Lambda 35?

The Perkin Elmer Lambda 35 is a UV/Vis spectrophotometer.[1]

Which series does the Lambda 35 belong to?

The Lambda 35 belongs to a series that includes the Lambda 25 and Lambda 45 models.[2]

Who manufactured the Perkin Elmer Lambda 35?

The Lambda 35 was manufactured by PerkinElmer.[5][1][2]

Not publicly documented

The following facts about the Lambda 35 are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the Lambda 35 are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • The control-system platform and OS era of the Lambda 35 are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the Lambda 35 are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the Lambda 35 is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

  • No publicly documented compatible parts, consumables, or accessories for the Lambda 35 are on record.

    Answerable by: an OEM parts catalog or a service engineer

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (5)Every fact above is drawn from these public sources
  1. [1]PerkinElmer Lambda 35 UV/Vis Spectrophotometer &#8211; Spectralab Scientific Inc. — spectralabsci.comspectralabsci.com
  2. [2]richmondscientific.com — richmondscientific.comrichmondscientific.com
  3. [3]ietltd.comietltd.com
  4. [4]leisales.comleisales.com