Agilent7700 Series ICP-MS
No photo available yet

Power

Single Phase, 200-240V, 50/60Hz, 30A[3]

Vacuum

Mechanical Pump[2]

Gas delivery

<2 cps[3]

What it is

General reference — not yet source-verified

Inductively coupled plasma mass spectrometers (ICP-MS) are analytical instruments designed for the detection and quantification of trace elements in liquid samples. The class combines an inductively coupled plasma source, which atomizes and ionizes the sample, with a mass spectrometer that separates ions based on their mass-to-charge ratio.

How it works

General reference — not yet source-verified

In an ICP-MS, the liquid sample is nebulized into a fine aerosol and carried into an argon plasma that reaches temperatures high enough to atomize and ionize most elements. The resulting ions are extracted through a sampling cone and focused by a series of ion lenses into the mass analyzer.

The mass analyzer, typically a quadrupole, separates ions by their mass-to-charge ratio. A detector counts the ions at each mass, producing a mass spectrum that shows the elemental composition of the sample. The entire process is computer-controlled, allowing automated analysis of multiple samples.

Applications

General reference — not yet source-verified

Common applications include trace metal analysis in environmental water and soil samples, quality control of ultrapure chemicals in semiconductor manufacturing, detection of heavy metals in food and pharmaceutical products, and isotopic ratio measurements in geochemistry.

What do the numbers mean?

Power & electrical2

Power Supply
Single Phase, 200-240V, 50/60Hz, 30A[3]
Accurate?
RF power range
500 W to 1600 W, in steps of 10 W[1]
Accurate?

Vacuum & pumping2

Mechanical Pump[2]
Accurate?
Argon Gas Supply (supply pressure)
500-700 kPa[3]
Accurate?

Gas & chemistry6

Background on-mass (No gas, 9 amu) - 7700x
<2 cps[3]
Accurate?
Background on-mass (No gas, 9 amu) - 7700s
<3 cps[3]
Accurate?
Background on-mass (He cell gas, 9 amu) - 7700x
<0.5 cps[3]
Accurate?
Background on-mass (He cell gas, 9 amu) - 7700s
<0.5 cps[3]
Accurate?
Argon Gas Supply (minimum purity)
99.99%[3]
Accurate?
Argon Gas Supply (maximum flow rate)
20 L/min[3]
Accurate?

Optics & imaging1

Detector
Electron Multiplier (pulse and analog modes)[4]
Accurate?

Control & software1

Computer Loaded with Software[2]
Accurate?

Dimensions & facilities5

Dimensions (mainframe, W x D x H)
730 x 620 x 595 mm[3]
Accurate?
Weight (mainframe)
115 kg[3]
Accurate?
Mainframe dimensions (W x D x H)
730 mm x 620 mm x 595 mm[3]
Accurate?
Mainframe weight
115 kg[3]
Accurate?
Sampling depth
3 to 28 mm[3]
Accurate?

Configuration & options26

Agilent 7700 Series ICP/MS[2]
Accurate?
Autosampler[2]
Accurate?
Chiller[2]
Accurate?
Windows OS, Monitor, Keyboard & Mouse[2]
Accurate?
Sold AS-IS[2]
Accurate?
Mass Analyzer
Quadrupole[3]
Accurate?
Mass Range
2-260 amu[4]2 sources
Accurate?
Scan Speed
100 milliseconds (for full mass range)[4]
Accurate?
Sensitivity (7700x) Li (7)
50 Mcps/ppm[3]
Accurate?
Sensitivity (7700x) Y (89)
160 Mcps/ppm[3]
Accurate?
Sensitivity (7700x) Tl (205)
80 Mcps/ppm[3]
Accurate?
Sensitivity (7700s) Li (7)
50 Mcps/ppm[3]
Accurate?
Sensitivity (7700s) Y (89)
240 Mcps/ppm[3]
Accurate?
Sensitivity (7700s) Tl (205)
120 Mcps/ppm[3]
Accurate?
Sensitivity (7700e) Li (7)
20 Mcps/ppm[1]
Accurate?
Sensitivity (7700e) Y (89)
80 Mcps/ppm[1]
Accurate?
Sensitivity (7700e) Tl (205)
40 Mcps/ppm[1]
Accurate?
Oxide ratio (CeO/Ce) - 7700x
<1.5%[3]
Accurate?
Oxide ratio (CeO/Ce) - 7700s
<3%[3]
Accurate?
Doubly-charged ratio (Ce2+/Ce) - 7700x
<3%[3]
Accurate?
Doubly-charged ratio (Ce2+/Ce) - 7700s
<6%[3]
Accurate?
Short-term stability (20 min, %RSD)
<3% (7700x, 7700s)[3]
Accurate?
Long-term stability (2 hours, %RSD)
<4% (7700x, 7700s)[3]
Accurate?
Sensitivity (Y 89)
160 Mcps/ppm (7700x) / 240 Mcps/ppm (7700s)[3]
Accurate?
Sensitivity (Tl 205)
80 Mcps/ppm (7700x) / 120 Mcps/ppm (7700s)[3]
Accurate?
Oxide ratio (CeO/Ce)
<1.5% (7700x) / <3% (7700s)[3]
Accurate?

Vintage & configurations

  1. 2026OEM support ended[1]

    Agilent's 10-year Value Promise guarantees at least 10 years of use from the date of purchase.

Documented models & variants

DesignationGenerationVintageChangesSource
7700x——The 7700x is the standard workhorse model, including High Matrix Introduction (HMI) and a He cell gas line for ORS3 as standard.ietltd.com[3]
7700s——The 7700s is configured for semiconductor performance, including a reaction gas line for H2, a fifth carrier gas line, a MicroFlow (PFA) nebulizer, and Pt interface cones.ietltd.com[3]
7700e——The 7700e is a streamlined version with a Core version of MassHunter software, lower sensitivity, and does not include HMI, option gas line, or H2 cell gas line as standard.antpedia.com[1]
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • ConfigurationMechanical Pump[2]
  • Background on-mass (No gas, 9 amu) - 7700x<2 cps[3]
  • Background on-mass (No gas, 9 amu) - 7700s<3 cps[3]
  • Background on-mass (He cell gas, 9 amu) - 7700x<0.5 cps[3]
  • Background on-mass (He cell gas, 9 amu) - 7700s<0.5 cps[3]
  • Power SupplySingle Phase, 200-240V, 50/60Hz, 30A[3]
  • Argon Gas Supply (minimum purity)99.99%[3]
  • Argon Gas Supply (maximum flow rate)20 L/min[3]
  • Argon Gas Supply (supply pressure)500-700 kPa[3]
  • RF power range500 W to 1600 W, in steps of 10 W[1]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What are the main models in the 7700 Series?

The 7700 Series comprises three models: the 7700x (routine workhorse), the 7700s (semiconductor-configured), and the 7700e (streamlined).[3]

Not publicly documented

The following facts about the 7700 Series ICP-MS are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • The control-system platform and OS era of the 7700 Series ICP-MS are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the 7700 Series ICP-MS are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the 7700 Series ICP-MS is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

  • No publicly documented compatible parts, consumables, or accessories for the 7700 Series ICP-MS are on record.

    Answerable by: an OEM parts catalog or a service engineer

  • No publicly hosted manuals, SOPs, or datasheets for the 7700 Series ICP-MS are on record.

    Answerable by: university cleanroom staff or an OEM application specialist

No research found yet — worked with this tool? Share what you know.

Sources & citations