Waferpedia

Comparison ledger

MA6versusMA100 / 150e Gen2

Karl Suss MA6, SUSS MicroTec MA100 / 150e Gen2, side by side. Every value shown is drawn from the cited encyclopedia entries.

Side-by-side comparison of selected equipment models
Attribute
MA6Karl Suss
MA100 / 150e Gen2SUSS MicroTec
OEMKarl SussSUSS MicroTec
CategoryLithographyLithography
Wafer size150mm150mm
Process node1 micron and better; 0.6 µm under optimum conditions in vacuum mode—
Introduced——
Production run——
Lifecycle——
Lifecycle milestones——
Control system——
Generationi-lineGen2
FamilyKarl Suss MA6SUSS MicroTec MA100 / 150e Gen2
Cited variants—
Specifications
Substrate range5 x 5 mm dies to wafers up to 150 mm[3]—
Maximum substrate thickness6 mm[3]—
Resolution0.6 µm under optimum conditions in vacuum mode[3]—
Light sourceBroadband Hg lamp[3]—
Wavelength channelsChannel 1 (405 nm) and Channel 2 (365 nm)[3]—
Wafer chucks available3, 4, 6 inch[3]—
Mask holders available4, 5, 7 inch[3]—
Top side alignment precision0.5 µm[3]—
Objectives5X, 10X, and 20X[3]—
Alignment featuresSplitfield microscope with storage and automatic movement between reference points; Bottom-Side Alignment (BSA); Wedge Error Compensation (WEC)[3]—
Exposure modesproximity, soft contact, hard contact, and vacuum[3]—
Alignment accuracy0.5 µm (motorized top side alignment)[3]—
Microscope objectives5X, 10X, and 20X[3]—
Wedge Error CompensationYes (ensures parallelism between substrate and photomask)[3]—
Maximum substrate size150 mm (wafers) or 5x5 mm to 150 mm irregular substrates[3]—
Resolution (vacuum mode)0.6 µm (optimum conditions)[3]—
Bottom-Side AlignmentAvailable (BSA)[3]—
Resolution (optimum vacuum mode)0.6 µm[3]—
Resolution (typical)1 µm and better[4]—
Exposure methodsflood, proximity, soft contact, hard contact, low vacuum, vacuum[4]—
Mask sizes2.5x2.5, 4x4, 5x5, 8x8 inches[4]—
Wafer size (top-side alignment)up to 6 inches (2, 3, 4, 6 inch)[4]—
Wafer size (bottom-side alignment)3 and 4 inch[4]—
Wavelength rangeUV broadband 250-450 nm; i-line 365 nm; g-line 436 nm[4]—
Light source (UIC)Broadband Hg lamp with Channel 1 (405 nm) and Channel 2 (365 nm)[3]—
Alignment accuracy (motorized top-side)0.5 µm[3]—
Type—Automated system[5]
Wafer size—up to 150 mm[5]
Function—Mask Alignment, Exposure[5]

Plan your fab

Building a Lithography process? Get a complete equipment list.

Open the fab equipment planner →
Sources (5)Every fact above is drawn from these public sources
  1. [1]ece.utoronto.caece.utoronto.ca
  2. [2]louisville.edulouisville.edu
  3. [3]ncf.uic.eduncf.uic.edu
  4. [4]nanolab.ucla.edunanolab.ucla.edu