Comparison ledger
Applied Materials 0010-21246 Part, ESPEC TSE 12 A Thermal Shock, side by side. Every value shown is drawn from the cited encyclopedia entries.
| Attribute | 0010-21246 PartApplied Materials | |
|---|---|---|
| OEM | Applied Materials | ESPEC |
| Category | — | — |
| Wafer size | — | — |
| Process node | — | — |
| Introduced | — | — |
| Production run | — | — |
| Lifecycle | — | — |
| Lifecycle milestones | — | — |
| Control system | — | — |
| Generation | — | — |
| Family | — | ESPEC TSE 12 A Thermal Shock |
| Specifications | ||
| Phase | — | 3 Phase[1] |
| Frequency | — | 60 Hz[1] |
| Interior Volume | — | 0.4 cu. ft. (11 Liters)[2] |
| Interior Dimensions (W x D x H) | — | 12.5" x 9.1" x 5.8" (320 x 230 x 150 mm)[2] |
| Exterior Dimensions (W x D x H) | — | 26" x 42" x 79" (NEC compliant version) (680 x 1050 x 2007 mm)[2] |
| Temperature Range | — | -65 to 200°C (-85 to 392°F)[2] |
| Power | — | 208V (±5%) 3ph 60Hz 30 Service Amps (26 FLA)[2] |
| Maximum Sound Level (1 meter away) | — | 60 dBA[2] |
| Temperature Fluctuation (at control sensor) | — | ±0.5°C[2] |
| Refrigeration System | — | Cascade 2hp Scroll[2] |
| Transfer Mechanism | — | Vertical-lift, direct-drive system for reliable transfer between zones[3] |
| Maximum Load | — | 17 lb. (4 kg)[3] |
| Specimen Baskets | — | Two specimen baskets[3] |
| Cable Port | — | Tube for cables to test load, 2" diameter[3] |
| Specimen Power Supply Terminal | — | Included[3] |
| Refrigeration Type | — | Non-CFC cascade refrigeration system using 2-hp compressors[3] |
| Air Circulation | — | Two sirocco blower motors circulate the air (one per zone)[3] |
| Controller | — | ESPEC's Exclusive P-310 Programmer with 7" color touch-screen[3] |
| Programs | — | 20 user-definable programs can be saved with alphanumeric names[3] |
| Connectivity | — | Ethernet with browser-based access is standard, plus USB[3] |
| Utility Requirement | — | The only utility required is 208V power[3] |
| Width | — | Only 26 inches wide[3] |
| Compliance | — | Meets Mil-Std 883 1010.7/1010.8 thermal cycling requirements with 4 kg load; also can be used for IEC 61300-2-47, JEDEC JESD47E, JESD22-A104, and others[3] |
| Test Performance Example | — | Test: Mil-Std 883 1010.7 Range: -65 to 150°C Recovery: 15 min. (worst-case on-product) Load: 6 lb of IC chips[2] |
| Test performance example | — | Mil-Std 883 1010.7; range -65 to 150°C; recovery 15 min. (worst-case on-product); load 6 lb of IC chips[2] |
| Heat-to-room peak | — | 17,000 BTU/Hr.[2] |
| Maximum sound level | — | 60 dBA (1 meter away)[2] |
| Temperature fluctuation | — | ±0.5°C at control sensor[2] |
| Maximum load | — | 17 lb.[2] |
| Cable tube diameter | — | 2 in.[2] |
| Display | — | 7 in. color touch-screen[2] |